Visual error signals
272
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
26.3
Visual error signals
Displays
Meaning
F
UNCTION
TEST
Error on the operating table. The last 10
messages can be displayed on the remote
control. Evaluation possible using the event log
(CAN test center). All messages are listed in
the text file.
Lights red
B
ATTERY
STATUS
+
E
XTERNAL
POWER
SUPPLY
F
UNCTION
TEST
Operating table cannot be switched off.
Communication with the power supply unit is
disrupted. The operating table must be
restarted using the reset switch.
Not illuminated
Not illuminated
Lights red
B
ATTERY
STATUS
+
E
XTERNAL
POWER
SUPPLY
F
UNCTION
TEST
Operating table cannot be switched off.
Communication with the battery is disrupted.
The operating table must be restarted using the
reset switch.
Flashes red
Not illuminated
Lights red
Содержание TruSystem 7000
Страница 1: ...TruSystem 7000 Operating Table Service Manual...
Страница 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Страница 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 305: ......