6-36
Time
M
easure
d
val
ue
Removal
value
Median
value
●
●
④
②
③
⑤
Target of movement time:5[s]
①
●
●
Removal
value
Measurement Support data
- Averaging Time
Mean movement time for measurement value
(1sec
to
120sec)
- Median Filter
Number of excluding deviations in the measurement values within the
mean movement time.
(0sec to 120sec / Default: 0 sec)
Fig. 6.6.2
Median Filter
Fig. 6.6.2 indicates setting at “Average Time”5sec and “Median Filter”2sec.
KRG-10 will generate 1 measured result (process value) every 1 second.
5 sec means KRG-10 can generate 5 measured values.
In case that you will set 2 sec for Median Filter value, it means removing 2 values from 5
measured values to get median value by 3 measured values.
Number
of
measurement
values
–
Median
Filter
time
Calculated
values
To get smooth value, Median Filter should be set as minimum.
On the contrary, Median Filter should be set as maximum to get vivid value such as rapid
level changing condition.
- Alarm Delay
Time until alarm is output after data was missed
(1sec
to
120sec)
- Re-search Delay
Time until searching the liquid level echo after data was missed
(1sec
to
120sec)
- Surface Detect Mode
Echo detected as the liquid level echo
(Max Echo / first Echo)
Fig. 6.6.3
Surface Detect Mode
Distance
S
ig
n
al
Amp
lit
ude
First Echo
Max Echo
Содержание KRG-10
Страница 1: ...Document No KF13 001K Microwave Level Gauge RG 10 User s Guide...
Страница 9: ...8 Blank...
Страница 13: ...12 Blank...
Страница 17: ...1 4 Fig 1 2 4 Transmitter head with rod antenna 1 Transmitter head 2 Flange option 3 1 rod antenna...
Страница 51: ...3 8 Blank...
Страница 55: ...4 4 Blank...
Страница 171: ...9 2 Blank...
Страница 181: ...10 10 10 4 Optional specifications Table 10 4 1 Optional specifications CD R Configuration software...
Страница 190: ...10 19 M20 1 5 Fig 10 7 6 Cable gland Only Non Ex type...
Страница 193: ...11 2 Blank...
Страница 197: ...12 4 Blank...
Страница 199: ...Blank...