3-4
<Hazardous area>
<Non-hazardous area>
+
-
KRG-10
Intrinsic
safety
barrier
Power
supply
(2) Ex ia connection example.
Operating voltage: 12 to 30V DC
Output signal: 4 to 20mA DC
Electrical parameters for the entity concept
Ui=30V
Ii=93mA
Pi=0.7W
Ci
≈
0
Li
≈
0
Fig. 3.1.5 Ex ia connection example.
NOTE:
・
Select appropriate ATEX / IECEx classified product to satisfy the Ex requirement.
・
Grounding must be connected.
・
Do not open either of lids, while the power is supplied or in the explosive gas and dust atmosphere.
・
Do not make static electricity.
・
Do not use paint thinner or other chemical products. Wipe away any dirt with a soft cloth.
・
Do not modify any of structure.
・
Following requirements must be kept in accordance with ATEX / IECEx regulation.
Ambient temperature
-40 to 60°C
Process temperature
Please refer to the “Safety Instructions”.
・
Intrinsic safety barrier must satisfy the rating of electrical parameters for the entity concept.
・
Cable inductance (=Lw) and capacitance (=Cw) should be chosen so as to lower the rating of
Intrinsic safety barrier allowable inductance (=Lo) and capacitance (=Lo).
・
Perform grounding structure.
CAUTION
Ensure the requirement of wiring with Ex product must be satisfied.
Содержание KRG-10
Страница 1: ...Document No KF13 001K Microwave Level Gauge RG 10 User s Guide...
Страница 9: ...8 Blank...
Страница 13: ...12 Blank...
Страница 17: ...1 4 Fig 1 2 4 Transmitter head with rod antenna 1 Transmitter head 2 Flange option 3 1 rod antenna...
Страница 51: ...3 8 Blank...
Страница 55: ...4 4 Blank...
Страница 171: ...9 2 Blank...
Страница 181: ...10 10 10 4 Optional specifications Table 10 4 1 Optional specifications CD R Configuration software...
Страница 190: ...10 19 M20 1 5 Fig 10 7 6 Cable gland Only Non Ex type...
Страница 193: ...11 2 Blank...
Страница 197: ...12 4 Blank...
Страница 199: ...Blank...