6-81
(1) Executing Parameter Initialization(*It is supported by firmware version between 1.4.1
and less than 1.5.0)
To initialize device parameters, select "Parameter Initialization" ( ), and then click the
"Execute" button. The following window will be displayed, so then click "OK" to initialize the
device parameters.
Fig. 6.10.2 Parameter initialization confirmation dialog box
The parameters that are initialized by config software
Parameters except the noise table, fixed disturbance echo, and HART bus address.
(2) Executing Parameter Backup(*It is supported by firmware version 1.5.0 or later)
To backup device parameters, select "Parameter Backup" ( ), and then click the "Execute"
button. The following window will be displayed, so then click "OK", the setting parameters are
stored in the inside of the device. When parameter backup is executed, you have to be very
careful because factory-configured parameters are overwritten and removed.
Fig. 6.10.3 Parameter backup confirmation dialog box
Please don't turn off power for five seconds after executing parameter backup because it leads
to device failure.
The parameters that are initialized by config software
Parameters except the noise table, fixed disturbance echo, and HART bus address.
(3) Executing Parameter Recovery(*It is supported by firmware version 1.5.0 or later)
To recovery device parameters, select "Parameter Recovery" ( ), and then click the
"Execute" button. The following window will be displayed, so then click "OK", the parameters
stored in the inside of the device are loaded. When parameter backup is never executed,
factory-configured parameters are loaded.
Содержание KRG-10
Страница 1: ...Document No KF13 001K Microwave Level Gauge RG 10 User s Guide...
Страница 9: ...8 Blank...
Страница 13: ...12 Blank...
Страница 17: ...1 4 Fig 1 2 4 Transmitter head with rod antenna 1 Transmitter head 2 Flange option 3 1 rod antenna...
Страница 51: ...3 8 Blank...
Страница 55: ...4 4 Blank...
Страница 171: ...9 2 Blank...
Страница 181: ...10 10 10 4 Optional specifications Table 10 4 1 Optional specifications CD R Configuration software...
Страница 190: ...10 19 M20 1 5 Fig 10 7 6 Cable gland Only Non Ex type...
Страница 193: ...11 2 Blank...
Страница 197: ...12 4 Blank...
Страница 199: ...Blank...