C-Link Protocol Commands
Diagnostics
B-16
Model 15
i
Instruction Manual
Thermo Fisher Scientific
set alarm trig conc
value
This command sets the HCl concentration alarm minimum
value
, where
value
is set to either floor or ceiling, according to
Table B–4
. The following
example sets the HCl concentration minimum alarm trigger to ceiling.
Send:
set alarm trig conc 1
Receive:
set alarm trig conc 1 ok
Table B–4.
Alarm Trigger Values
Value
Alarm Trigger
00 Floor
01 Ceiling
agc int
This command reports the current intensity of the reference channel AGC
circuit. The following example reports that the current AGC intensity is
250,000 Hz.
Send:
agc
int
Receive:
agc int 250000 Hz
bias voltage
This command reports the current IR bias supply voltage. The following
example reports that the bias voltage is -102.3 volts
Send:
bias
voltage
Receive:
bias voltage -102.300 V
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
mb
Receive:
diag volt mb 24.10 14.90 4.90 3.20 -3.20
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18
MOT, and VBIAS. Each voltage value is separated by a space.
Send:
diag
volt
mib
Receive:
diag volt mib 23.98 14.80 -14.90 4.96 3.20 18.00 17.75
-108.30
Diagnostics