C-Link Protocol Commands
Measurements
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-9
Command
Description
Page
up
Simulates pressing up pushbutton
B-27
zero
Sets zero/span valves to zero mode
B-30
avg time
high avg time
low avg time
These commands report the averaging time in seconds when operating in
single range, or averaging time used with the high and low ranges when
operating in dual or auto range mode. The following example shows that
the averaging time is 300 seconds, according to
Table B–3
.
Send:
avg
time
Receive:
avg time 11:300 sec
set avg time
selection
set high avg time
selection
set low avg time
selection
These commands set the averaging time, high and low averaging times,
according to
Table B–3
. The following example sets the averaging time to
120 seconds.
Send:
set avg time 8
Receive:
set avg time 8 ok
Table B–3.
Averaging Times
Selection
Averaging Time
0 1
seconds
1 2
2 5
3 10
4 20
5 30
6 60
7 90
8 120
9 180
10 240
11 300
Measurements