Operation
Alarms Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-65
●
In the Main Menu, choose Alarms > Sample Flow >
Min
or
Max
.
SAMPLE FLOW:
ACTUAL MIN: 0.350 LPM
SET MIN TO: 0.500 LPM ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Bias Voltage screen displays the current bias voltage reading and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from -130 to -100 volts. If the bias voltage reading goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Bias Voltage
.
BIAS VOLTAGE:
ACTUAL -110 V
>MIN -130 V
MAX -100 V
RANGE AVG DIAGS ALARM
The Minimum Bias Voltage alarm limit screen is used to change the
minimum bias voltage alarm limit. The minimum and maximum bias
voltage screens function the same way.
●
In the Main Menu, choose Alarms > Bias Voltage >
Min
or
Max
.
BIAS VOLTAGE:
ACTUAL MIN: -130 V
SET MIN TO: -120 V ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The AGC Intensity screen displays the current AGC Intensity reading and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 150,000 to 300,000 Hz. If the AGC intensity reading goes
Bias Voltage
Min and Max Bias
Voltage Limits
AGC Intensity