Installation
Setup Procedure
Thermo Fisher Scientific
Model 15
i
Instruction Manual
2-3
5.
Check for possible damage during shipment.
6.
Check that all connectors and circuit boards are firmly attached.
7.
Re-install the cover.
8.
Remove any protective plastic material from the case exterior.
Use the following procedure to setup the instrument:
1.
Connect the sample line to the SAMPLE bulkhead on the rear panel
(
Figure 2–3
). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of
FEP Teflon, 316 stainless steel, borosilicate glass, or similar tubing with
an OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the
tubing should be less than 10 feet.
Figure 2–3.
Model 15
i
Rear Panel
Note
Gas must be delivered to the instrument free of particulates. It may
be necessary to use the Teflon particulate filter as described in “
Teflon
Particulate Filter
” on page
9-2
.
▲
Setup Procedure