C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-15
alarm pressure min
alarm pressure max
These commands report the current pressure alarm minimum and
maximum value settings. The following example reports that the pressure
alarm minimum value is 765.9 mmHg.
Send:
alarm pressure min
Receive:
alarm pressure min 765.0, mm Hg, actual 765.9
set alarm pressure min
value
set alarm pressure max
value
These commands set the pressure alarm minimum and maximum values to
value
, where
value
is a floating-point number representing pressure alarm
limits in millimeters of mercury. The following example sets the pressure
alarm maximum value to 900 mmHg.
Send:
set alarm pressure max 900
Receive:
set alarm pressure max 900 ok
alarm sample flow min
alarm sample flow max
These commands report the current sample flow alarm minimum and
maximum value settings. The following example reports that the sample
flow alarm minimum value is 2 LPM.
Send:
alarm sample flow min
Receive:
alarm sample flow min 2.000 l/min
set alarm sample flow min
value
set alarm sample flow max
value
These commands set the sample flow alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing sample
flow alarm limits in liters per minute. The following example sets the
sample flow alarm maximum value to 1 LPM.
Send:
set alarm sample flow max 1
Receive:
set alarm sample flow max 1 ok
alarm trig conc
This command reports the HCl concentration alarm trigger action for
minimum alarm to either floor or ceiling. The following example below
shows the HCl concentration minimum alarm trigger to ceiling, according
to
Table B–4
.
Send:
alarm trig conc
Receive:
alarm trig conc 1