Servicing
Measurement Interface Board Replacement
7-38
Model 15
i
Instruction Manual
Thermo Fisher Scientific
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the I/O expansion board (optional), if used. See the “
I/O
Expansion Board (Optional) Replacement
” procedure in this chapter.
3.
Remove the digital output board. See the “
Digital Output Board
Replacement
” procedure in this chapter.
4.
Unplug all connectors from the motherboard. Note connector locations
to facilitate re-connection.
5.
Using the nut driver, remove the six standoffs securing the board to the
rear panel (
Figure 7–17
).
6.
Pop the motherboard off of the support bracket, and remove the
motherboard.
7.
To install the motherboard, follow the previous steps in reverse.
8.
Calibrate the analog voltage outputs as defined earlier in this chapter
(all ranges).
Use the following procedure to replace the measurement interface board
(
Figure 7–18
).
Equipment Required:
Measurement interface board
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
Measurement
Interface Board
Replacement