Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
Thermo Fisher Scientific
Model 15
i
Instruction Manual
D-7
Table D–1.
Operating Status for Model 15
i
D7 D6 D5 D4 D3 D2 D1 D0
Æ
Bit
8
7
6
5
4
3
2
1
Æ
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Operating status:
Service
Mode
(On)
0 0 0 0
0 0 0 1
Maintenance
(Local)
0 0 0 0
0 0 1 0
Zero
Gas
(On)
0 0 0 0
0 1 0 0
Span
Gas
(On)
0 0 0 0
1 0 0 0
Gas Unit Indication (ppm or ppb)
0
0
0
1
0
0
0
0
Ozonator
(Off)
0 0 1 0
0 0 0 0
PMT
(Off)
0 1 0 0
0 0 0 0
Not
Used
1 0 0 0
0 0 0 0
Table D–2.
Error Status for Model 15
i
D7 D6 D5 D4 D3 D2 D1 D0
Æ
Bit
8
7
6
5
4
3
2
1
Æ
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Error status:
Internal
Temperature
Alarm
0 0 0 0
0 0 0 1
Optical Chamber Temperature
Alarm
0 0 0 0
0 0 1 0
AGC
Intensity
Alarm
0 0 0 0
0 1 0 0
Bias
Voltage
Alarm
0 0 0 0
1 0 0 0
Pressure
Alarm
0 0 0 1
0 0 0 0
Sample
Flow
Alarm
0 0 1 0
0 0 0 0
Motor
Speed
Alarm
0 1 0 0
0 0 0 0
Not
Used
1 0 0 0
0 0 0 0