Servicing
Flow Transducer Calibration
Thermo Fisher Scientific
Model 15
i
Instruction Manual
7-25
Figure 7–12.
Replacing the Flow Transducer
5.
Install the new flow transducer following the previous steps in reverse.
6.
Calibrate the flow transducer. Refer to the “
Flow Transducer
Calibration
” procedure that follows.
Use the following procedure to calibrate the flow transducer.
Equipment Required:
Calibrated flow sensor
WARNING
The service procedures in this manual are restricted to qualified
service representatives.
▲
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Disconnect the pump cable from AC PUMP connector on the
measurement interface board.
Flow Transducer
Calibration
Retaining Screws (2)