Operation
Diagnostics Menu
3-58
Model 15
i
Instruction Manual
Thermo Fisher Scientific
AGC INTENSITY:
INTENSITY: 250000 Hz
RANGE AVG DIAGS ALARM
The Motor Speed screen (read only) displays the status of the chopper
motor. A reading of 100.0% means that the motor speed is correct. A
reading other than 100.0% indicates that there is a problem with the
chopper motor or power supply.
●
In the Main Menu, choose Diagnostics >
Motor Speed
.
MOTOR SPEED:
SPEED: 100.00 %
RANGE AVG DIAGS ALARM
The Analog Input Readings screen (read only) displays the 8 user-scaled
analog readings (if the I/O expansion board option is installed).
●
In the Main Menu, choose Diagnostics >
Analog Input Readings
.
ANALOG INPUT READINGS:
>HCL 10.2 PPM
SO2 18.2 PPM
CO2 250 PPM
FL1 20.42 LPM
IO5 0.0 V
ª
RANGE AVG DIAGS ALARM
The Analog Input Voltages screen (read only) displays the 8 raw analog
voltage readings (if the I/O expansion board option is installed).
●
In the Main Menu, choose Diagnostics >
Analog Input Voltages
.
Motor Speed
Analog Input
Readings
Analog Input
Voltages