Operation
Service Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-73
●
In the Main Menu, choose Service > Flow Calibration >
Span
.
CALIBRATE FLOW SPAN:
CURRENTLY: 1.139 LPM
SET TO: 1.14
1
LPM ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Restore Default Flow Calibration screen allows the user to reset the
flow calibration configuration values to factory defaults.
●
In the Main Menu, choose Service > Flow Calibration >
Set Defaults
.
The Initial Sample/Reference (S/R) Ratio screen displays both the initial
S/R ratio and the current S/R ratio. The initial S/R ratio is determined at
the factory, and is used to correct for the slight variations found from one
correlation wheel to another. The only time the initial S/R ratio should be
changed is when the correlation wheel is replaced or sample/reference is
between 1.05 and 1.15. The initial sample/reference ratio screen is visible
only when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Initial S/R Ratio
.
Restore Default Flow
Calibration
Initial
Sample/Reference
Ratio
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM