Operation
Instrument Controls Menu
3-42
Model 15
i
Instruction Manual
Thermo Fisher Scientific
The Non-Alarm screen allows the user to select a non-alarm status for the
selected relay output. The currently selected item is shown by “<--” after it.
The zero, span, and sample modes are only present if the zero/span valve
option is installed.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > select Relay > Instrument State >
Non-Alarm
.
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE
LOCAL/REMOTE
SERVICE
UNITS
ª
RANGE AVG DIAGS ALARM
ZERO MODE
SPAN MODE
SAMPLE MODE
PURGE MODE
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital inputs must be asserted for at least one second for the
action to be activated.
▲
Not all of the I/O available in the instrument are brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. (See optional
Terminal Block and Cable Kits
.)
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Setting
s
.
DIGITAL INPUT SETTINGS:
>1 NOP SET BACKGROUND
2 NOP CAL TO LO SPAN
3 NOP AOUTS TO ZERO
4 NOP AOUTS TO FS
5 NOP NONE
ª
RANGE AVG DIAGS ALARM
Non-Alarm
Digital Input Settings