C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-29
Send:
range
Receive:
range 3: 5.000E+04 ppb
set range
selection
set high range
selection
set low range
selection
These commands select the HCl full-scale ranges, according to
Table B–7
.
The following example sets the HCl full-scale range to 2000 ppm.
Send:
set range 8
Receive:
set range 8 ok
Table B–7.
Ranges
Selection
ppm
mg/m
3
0 5
10
1 10
20
2 20
50
3 50
100
4 100
200
5 200
500
6 500
1000
7 1000
2000
8 2000
5000
9 5000
7500
10 C1
C1
11 C2
C2
12 C3
C3
custom
range
range
= | 1 | 2 | 3 |
This command reports the user-defined value of custom
range
1, 2, or 3.
The following example reports that custom range 1 is defined to 55 ppm.
Send:
custom
1
Receive:
custom 1 5.500E+04 ppb
set custom
range
range
value
set custom 1
value
set custom 2
value
set custom 3
value