Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-41
The Instrument State submenu allows the user to select the instrument
state that is assigned to the selected relay output. A submenu lists signal
types of either alarms or non-alarm to choose from.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > select Relay >
Instrument State
.
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
RANGE AVG DIAGS ALARM
The Alarms screen allows the user to select an alarm status for the selected
relay output. The currently selected item is shown by “<--” after it. The
I/O board status alarm is only present if the I/O expansion board is
installed. Zero and Span Check Cal are only present if the automated
zero/span check is enabled. See the Alarms Menu section later in this
chapter for more information on alarm items.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > select Relay > Instrument State >
Alarms
.
ALARM STATUS ITEMS:
>NONE
GEN ALARM <--
CONC MAX
CONC MIN
INT TEMP
ª
RANGE AVG DIAGS ALARM
BENCH TEMP
PRESSURE
SAMPLE FLOW
INTENSITY
MOTOR SPEED
BIAS VOLTAGE
MB STATUS
MIB STATUS
I/O BD STATUS
CONC ALARM
ZERO CHK/CAL
SPAN CHK/CAL
Instrument State
Alarms