Operation
Service Menu
3-78
Model 15
i
Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Service > Analog Out Cal > select Channel
>
Calibrate Full Scale
.
ANALOG OUTPUT CAL: SPAN
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 3397
SAVE VALUE
©ª
INC/DEC
SET OUTPUT TO: 10 V
RANGE AVG DIAGS ALARM
The Analog Input Calibration menu is used to calibrate the 8 analog input
channels at both zero and full-scale. The analog input calibration menu is
visible only when the I/O expansion board is installed and when the
instrument is in service mode. For more information on the service mode,
see “
Service Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Analog Input Cal
.
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
ª
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
●
In the Main Menu, choose Service > Analog Input Cal > select Channel
>
Calibrate Zero
.
Analog Input Calibration
Analog Input Calibrate
Zero