The Flow Calibration submenu is used to calibrate the flow sensor to zero,
span, or restore factory default values. The flow calibration screen is visible
only when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Flow Calibration
.
FLOW SENSOR CAL:
>ZERO 121
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
The Calibrate Flow Zero screen calibrates the flow sensor at zero flow.
Note
The pump must be disconnected before performing the zero
calibration.
▲
●
In the Main Menu, choose Service > Flow Calibration >
Zero
.
CALIBRATE FLOW ZERO:
CURRENTLY: 1.139 LPM
SET TO: 0.000 LPM ?
DISCONNECT PUMP AND
SAVE CURRENT FLOW
RANGE AVG DIAGS ALARM
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibrate span point.
Note
An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
▲
Flow Calibration
Calibrate Flow Zero
Calibrate Flow Span