Operation
Instrument Controls Menu
3-34
Model 15
i
Instruction Manual
Thermo Fisher Scientific
INSTRUMENT ID:
CURRENTLY: 15
SET TO: 16 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Communication Protocol screen is used to change the protocol for
serial communications. Possible choices include: C-Link, MODBUS,
Geysitech, and Streaming Data.
●
In the Main Menu, choose Instrument Controls > Communication
Settings >
Communication Protocol
.
COMMUNICATION PROTOCOL:
CURRENTLY: CLINK
SET TO: STREAMING ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Streaming Data Configuration menu is used to allow for configuration
of the 8 streaming data output items, streaming interval, current data
format, and current timestamp setting. The Choose Stream Data submenu
displays a list of the analog output signal group choices to choose from.
Choices are Concentrations, Other Measurements, and Analog Inputs (if
the I/O expansion board option is installed).
●
In the Main Menu, choose Instrument Controls > Communication
Settings >
Streaming Data Config
.
STREAMING DATA CONFIG
>INTERVAL 10 SEC
ADD LABELS NO
PREPEND TIMESTAMP YES
ADD FLAGS YES
ITEM 1 HCL
ª
RANGE AVG DIAGS ALARM
The Streaming Data Interval screen is used to adjust how frequently a new
record will be generated. The following interval times are available: 1, 2, 5,
10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds.
Communication Protocol
Streaming Data Configuration
Streaming Data Interval