C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-19
in ASCII format with text. For details on how to decode the flag fields
within these records, see
Figure B–1
.
Send:
lr01
Receive: lr01
09:03 08-02-07 flags 8c060000 hcl 7349E+000 hihcl
5994E+000 intt 33.689 cht 44.484 pres 758.886 smplfl
1.085 speed 100.000 biasv -115.883 intensity 199940.000
erec
This command returns a snapshot of the main operating conditions
(measurements and status) at the time the command is issued. The
following example shows a typical response.
The format is defined by the current settings of the "format" and "erec
format" commands. For details on erec formatting, see the “
Record Layout
Definition
” section at the end of this appendix. For details on how to
decode the flag fields within these records, see the "flags" command.
Send:
erec
Receive: erec
09:01 08-02-07 flags 8C040510 hcl 7349E+000 4 hihcl -
50000.000 4 s/r 0.000 los/r 0.902 biasv -112.668 agci
196543.000 intt 34.023 cht 47.995 smpfl 1.085 pres
758.886 avgt1 10 avgt2 10 hclbkg -136.832 hclcoef 1.000
lohclcoef 1.000 hclrange 5000000.000 locorange
50000.000 motor 100.000
lrec
srec
lrec
xxxx yy
srec
xxxx yy
lrec
aa:bb oo-pp-qq yy
srec
aa:bb oo-pp-qq yy
xxxx
= the number of past records
yy
= the number of records to return (1 to 10)
aa
= hours (01 to 23)
bb
= minutes (01 to 59)
oo
= month (01 to 12)
pp
= day (01 to 31)
= year
These commands output lrecs or srecs and erec data. The output format is
determined by the “set lrec format”, and “set srec format” commands. The
logging time is determined by the “set lrec per” and “set srec per”
commands.