Operation
Service Menu
3-74
Model 15
i
Instruction Manual
Thermo Fisher Scientific
INITIAL S/R RATIO:
CURRENTLY: 1.100000
MEASURED: 1.100000
SET TO: 1.100001 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Up to three gas concentrations (cal-points) for each range may be calibrated
to using the following steps. Three cal-points will give the most accurate
readings over the entire range. The calibration process is sequential and will
work properly if all steps are followed in order. The following example
shows the multi-point calibration screen in single range mode. In dual or
auto range modes, “HI” or “LO” is displayed to indicate the calibration of
the high or low concentrations and function the same way. For more
information on “
HI and LO Multi-Point Calibration
”, see Chapter 4,
“
Calibration
”. The multi-point calibration screen is visible only when the
instrument is in service mode. For more information on the service mode,
see “
Service Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Multipoint Calibration
.
MULTI-POINT CAL:
>CALIBRATE POINT 1
CALIBRATE POINT 2
CALIBRATE POINT 3
COEFFICIENTS
DEFAULT COEFFICIENTS
RANGE AVG DIAGS ALARM
The Calibrate Point 1 screen allows the user to view and set the selected
calibration point. The calibrate 2 and calibrate 3 screens function the same
way.
●
In the Main Menu, choose Service > Multipoint Cal >
Calibrate 1, 2
or
3
.
Multi-Point
Calibration
Calibrate Point 1/2/3