Calibration
Calibration
Thermo Fisher Scientific
Model 15
i
Instruction Manual
4-3
Use the following procedure to calibrate the instrument.
Connect the instrument and the calibration equipment as shown in
Figure
4–1
. If an optional sample line filter is used, the calibration must be
performed through this filter. Ensure that the flow rate into the output
manifold is greater than the total flow required by the analyzer and any
other flow demand connected to the manifold.
Figure 4–1.
Calibration Flow Schematic
Use the following procedure to set the HCl reading to zero.
1.
Allow sufficient time for the analyzer to warm up and stabilize.
2.
Adjust the dilution system of
Figure 4–1
so that zero air alone is
present in the manifold.
Since not all flow controllers have a positive shut off, it might be
necessary to disconnect the HCl input line and cap it.
3.
Allow the instrument to sample zero air until a stable reading is
obtained.
Calibration
Connect the Instrument
Zero Adjust
Содержание 15i