C-Link Protocol Commands
Alarms
B-12
Model 15
i
Instruction Manual
Thermo Fisher Scientific
Figure B–1.
Flags Field
alarm agc intensity min
alarm agc intensity max
These commands report the current agc intensity alarm minimum and
maximum value settings. The following example reports that the agc
intensity alarm minimum value is 150.
Send:
alarm agc intensity min
Receive:
alarm agc intensity min 150000.0 V
set alarm agc intensity min
value
set alarm agc intensity max
value
These commands set the agc intensity alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing agc
intensity alarm limits. The following example sets the agc intensity alarm
maximum value to 250000.
Send:
set alarm agc intensity max 250000
Receive:
set alarm agc intensity max 250000 ok
Alarms