Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
D-6
Model 15
i
Instruction Manual
Thermo Fisher Scientific
Example:
Geysitech Protocol with transmission of three concentrations (Instrument
ID is 1, Operation Status is 03, Error Status is 04):
Data Query String:
<STX>DA<CR>
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>0000000000 <SP>002 <SP>
↑
↑
↑
Address First Concentration(E-format)=25.78 1
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
↑
↑
↑
Second Concentration = 5.681 2 Third Concentration=11.75
0000000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response, if not valid in a particular
mode, will report a value of 0.0.
The following measurements reported in response to DA command are for
the Model 15
i.
The 1 measurement reported in single range mode includes:
●
HCl
The 2 measurements reported in dual or auto range modes include:
●
low HCl
●
high HCl
See
Table D–1
for operating status and
Table D–2
for error status for the
Model 15
i
.
Measurements reported in
response to DA command
Single Range Mode
Dual/Auto Range Mode
Operating and Error
Status