C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-21
Table B–5.
Record Output Formats
Format
Output Format
0
ASCII no text
1
ASCII with text
2 Binary
data
lrec layout
srec layout
erec layout
These commands report the layout (string indicating the data formats) for
data that is sent out in response to the lrec, srec, erec, and related
commands. For details on how to interpret the strings, see “Record Layout
Definition” later in this appendix.
Send:
lrec
layout
Receive:
lrec layout %s %s %lx %f %f %f %f %f %f %f %f %f
t
D
L
fffffffff
flags hcl hihcl intt cht pres smplfl speed biasv
intensity
lrec mem size
srec mem size
These commands report the number of lrecs and srecs that can be stored
with the current settings and the number of blocks reserved for lrecs and
srecs. The following example shows that 1075 blocks were reserved for lrecs
and the maximum number of lrecs that can be stored in memory is
241979. Memory allocation can be changed using the “malloc” command.
Send:
lrec mem size
Receive:
lrec mem size 241979 recs, 1075 blocks
lrec per
srec per
These commands report the lrecs and srecs logging period. The following
example shows that the srec logging period is 5 minutes.
Send:
srec
per
Receive:
srec per 5 min
set lrec per
value
set srec per
value
value
= | 1 | 5 | 15 | 30 | 60 |
These commands set the lrecs and srecs logging period to
value
in minutes.
The following example sets the lrec logging period to 15 minutes.