Operation
Service Menu
3-76
Model 15
i
Instruction Manual
Thermo Fisher Scientific
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Preamp Calibration
.
PREAMP BOARD CAL:
SAMPLE: 127807 Hz
REFERENCE: 148256 Hz
SET TO: 128 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Temperature Calibration screen allows the user to view and set the
ambient temperature sensor offset, which basically adjusts the temperature
sensor so that it reads correctly. Wait at least 30 seconds for the reading to
stabilize before saving the value. The temperature calibration screen is
visible only when the instrument is in service mode. For more information
on the service mode, see “
Service Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Temperature Calibration
.
CALIBRATE AMBIENT TEMP:
CURRENTLY: 32.3
o
C
SET TO: 032.
5
o
C ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see “
Service Mode
” earlier in
this chapter.
Temperature Calibration
Analog Output Calibration