Operation
Diagnostics Menu
3-60
Model 15
i
Instruction Manual
Thermo Fisher Scientific
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
RANGE AVG DIAGS ALARM
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and reset returns the
output to normal operation. The analog outputs are returned to normal
operation upon exiting this screen. The following example shows the screen
when all analog outputs are set to “normal” operating mode.
●
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1–6
,
or
Current Channel 1–6
.
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
RANGE AVG DIAGS ALARM
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
Note
If the analyzer is in service mode, pressing
will toggle the
selected item between yes and no. Zero/span valves needs to be set to ON
in order to do zero and span checks, and auto calibration needs to be set to
ON in order to do zero and span calibration resets.
▲
●
In the Main Menu, choose Diagnostics >
Instrument Configuration
.
Set Analog Outputs
Instrument Configuration