![Thermo 15i Скачать руководство пользователя страница 111](http://html.mh-extra.com/html/thermo/15i/15i_instruction-manual_1104215111.webp)
Operation
Service Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-75
SET CAL GAS POINT:
RANGE: 5000
CAL FACTOR 1: 1.0000
CAL GAS 1: 0040.00
0
?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Coefficients screen allows the user to view and re-calculate the
calibration coefficients.
●
In the Main Menu, choose Service > Multipoint Cal > select Cal Point
>
Coefficients
.
COEFFICIENTS:
1: 1.000000
2: 1.000000
3: 1.000000
RECALCULATE?
RANGE AVG DIAGS ALARM
The Default Coefficients screen allows the user to view and reset the
calibration coefficients to default values.
●
In the Main Menu, choose Service > Multipoint Cal > select Cal Point
>
Coefficients
.
COEFFICIENTS:
1: 1.000000
2: 1.000000
3: 1.000000
SET TO DEFAULTS?
RANGE AVG DIAGS ALARM
The Pre-amp Board Calibration screen is used to adjust the preamp board
calibration parameters. The preamp board calibration screen is visible only
when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in this chapter.
Coefficients
Default Coefficients
Preamp Board Calibration