71
Other Operation Procedures
Increases value
Dial
Decreases
value
9)
Turn the Dial to adjust the dosage.
Flow rate (mL/h) Dosage rate
Notes
• The range of the dosage rate is as follows:
Dose unit = mL/h :
0.01 to 150.00 mL/h (when using syringe of 5 mL)
0.01 to 300.00 mL/h (when using syringe of 10 mL, 20 mL, 30 mL)
0.01 to 1200.00 mL/h (when using syringe of 50/60 mL)
Dose unit = other than mL/h : 0.01 to 999.00
Follow the specified range if the flow rate range is specified for the drug library
registration. The available ranges of two types are shown below. (For details about
how to set, see the dedicated editing software (TERUFUSION Drug Library Manager
or TERUFUSION Software Package).)
- Soft limit
If the flow rate reaches the soft limit of the lower limit or the soft limit of the
upper limit, the “Soft lower limit reached, Override?” or “Soft upper limit
reached, Override?” confirmation message is displayed on the LCD. Press the
Confirmation switch to override the upper or lower limit. When the upper or lower
limit is overridden, the flow rate value turns orange.
- Hard limit
If the flow rate reaches the hard limit of the lower limit or the hard limit of the
upper limit, the “This is the hard lower limit” or “This is the hard upper limit”
confirmation message is displayed on the LCD. Even if the Confirmation switch
is pressed, the upper or lower limit cannot be overridden.
• When a dose unit other than mL/h is selected, the flow rate (mL/h) is automatically
calculated from the specified amount, solution, weight and dosage rate, and
displayed in the flow rate display.
Caution
• If the calculation of flow rate (mL/h) is outside of the setting range*,
“0.00 mL/h” or “-----.-- mL/h” is displayed and solution delivery does not start
even if the Start switch is pressed. Make the setting again within the flow rate
setting range.
* For details about the flow rate setting range, see [Notes] above or
“Specifications” on page 146.