Parts Description
Sensor icon display
Tab display
Volume delivered display
Status icon display
Flow rate display
Start up icon
Ready for bolus icon
Status icon display
Battery icon
Displays the remaining battery
level.
(lit)
When AC power is connected and charge is full.
and the remaining battery
level flashing alternately
When AC power is connected and battery is charging.
(flashing every other second)
Battery failure
(flashing every half second)
Battery and sub-battery* failure
and the remaining battery
level flashing alternately every
other second
Sub-battery failure or low battery
Battery charge level (displayed in 5 bars)
5 bars (green): Approx. 8 hours
4 bars (green): Approx. 6 hours
3 bars (green): Approx. 4.5 hours
2 bars (green): Approx. 3 hours
1 bar (orange): Approx. 30 min (when the Battery alarm has not been issued)
(In the case of a continuous solution delivery at a flow rate of 5 mL/h at temperature of 25°C,
with new battery after charging for 8 hours or more with the power turned off.)
The battery can be charged to approx. 80% in 3 hours.
* The sub-battery is used for a Power Failure alarm.
AC icon
Displays connection/
disconnection of the AC power
supply.
When AC power is not connected
When AC power is connected
Occlusion icon
The occlusion detection pressure
is displayed in 10 levels (1 to 10)
in the upper right of the icon.
The occlusion detection pressure level is 1.
The occlusion detection pressure level is 6 (Default).
The occlusion detection pressure level is 10.