79
See pages 15 and 16 for the menu list.
Notes
• While changing the menu settings, pressing the Start switch does not start solution delivery.
• Occlusion detection pressure, keypad lock, bolus, advanced dose mode and combination dose mode on Menu 1 can be selected
or set during solution delivery. Other items are not available.
• How to exit menus
Method 1) Press and hold the Display select switch in the menu for 2 seconds or more.
If the setting has not been completed, the setting is cancelled and the display goes to the flow rate screen.
Method 2) Press the Back/Mute switch repeatedly.
If the setting has not been completed, the setting is cancelled and the display goes back to the previous screen.
• If an alarm occurs before pressing the Confirmation switch or the Back/Mute switch while setting the menu, any changes will
return to the settings before the alarm was issued.
• Time before the screen returns to the flow rate screen.
(1) Level in which menu tab has been selected:
Goes back to the flow rate screen after 10 seconds of no operation.
(2) Levels on and after which menu tab has been specified:
Goes back to the flow rate screen after 2 minutes of no operation.
Changing Settings on the Menu