25
Prior to the initial use of the pump, set the date and time to record history correctly. (See page 110.)
Before use, carefully read “Precautions” and the labels attached to the product.
Prior to the Pump being Used
Connect to the AC power supply and charge (8 hours or more) with the power turned off.
Prior-to-use Inspection
Check the following points before use.
1)
There is no damage to the main unit, AC power cable or pole clamp, and that there is no
malfunction due to adhered drug solution.
Main unit/AC power cable
1
2
2
4
3
6
3
5
1. There is no damage, dent or adhered drug solution on the AC inlet and AC power cable.
2. There is no damage to the handle, rubber protector, tube holder and infrared communication window.
3. There is no damage or adhered drug solution on the slider, slider hook and flange holder (including Slit).
4. When moving the slider left and right with the clutch pushed, it should move smoothly, and when it is moved to the left
most position while the power is on, the Nearly Empty alarm should be issued. (See page 142.)
5. There is no adhered drug solution on the syringe clamp and that it moves back and forth smoothly.
6. There is no drug solution on the Dial, it turns smoothly, and flow rate setting can be set when the power is on.
If there is adhered drug solution, promptly clean in accordance with the directions on page 129.
Preparation