78
Other Operation Procedures
Caution
• If the calculation of flow rate (mL/h) is outside of the setting range*,
“0.00 mL/h” or “-----.-- mL/h” is displayed and solution delivery does not start
even if the Start switch is pressed. Make the setting again within the flow rate
setting range.
* For details about the flow rate setting range, see page 146 .
Start switch
Operation indicator (green flashing)
9)
Press the Start switch to start solution delivery.
Checks
• The set value is correct.
• The operation indicator flashes green.
• The volume delivered increases.
Notes
Calculation Formula of Flow Rate
Amount (mg) X 1000
(
*
2)
'RVHUDWHȝJNJPLQ;%RG\ZHLJKWNJ;6ROXWLRQP/
Amount (mg)
'RVHUDWHPJNJK
X
%RG\ZHLJKWNJ
X
6ROXWLRQP/
)ORZUDWHP/K
X 60
(
*
1)
*
1
: Coefficient to convert “minute (min.)” to “hour (h)”
*
2
: Coefficient to convert “mg” to “
μ
g”
• Switching the dose unit or re-selecting the same dose unit changes the values of flow rate, dosage, amount, solution, and
volume delivered to “0”.
• Even after the setting is completed, the TCI mode select screen appears when the power is turned off and on again.
• Patient name, Patient ID, Gender, Age, Height and Weight can be acquired via communication with an external device. For the
setting method, contact TERUMO trained service technicians.