Troubleshooting
Syringe Empty
Occurrence
Cause
Action
• LCD: The Syringe Empty
display flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
• The syringe is fully pressed.
• Replace the syringe if you wish to continue the infusion.
(See page 62.) The time from full-press to alarm
sounding varies according to the product used. The
table below shows an example.
* When new TERUMO syringes are used.
5 mL
syringe
10 mL
syringe
20 mL
syringe
30 mL
syringe
50/60 mL
syringe
1 mL/h Approx.
3 min.
Approx.
11 min.
Approx.
30 min.
Approx.
41 min.
Approx.
64 min.
5 mL/h Approx.
1.0 min.
Approx.
2.5 min.
Approx.
5.0 min.
Approx.
8.5 min.
Approx.
12.0 min.
Completion alarm
Occurrence
Cause
Action
• LCD: The Completion alarm
display flashes red.
• Operation indicator: Flashes
green.
• Buzzer: Activates.
• Pump: Switches to the keep
vein open (KVO) function and
continues solution delivery.
• The volume delivered has
reached the VTBI.
• A normal operation when the VTBI is set.
• Press the Stop switch to stop the KVO solution delivery.
(See page 57.)