155
Device Characteristics
Occlusion Characteristics
Typical occlusion detection ability data are occlusion pressure, Occlusion alarm occurrence time, and post-occlusion removal
bolus volume. The data below is a representative example measured for flow rates of 1 mL/h and 5 mL/h, with a three-stage
occlusion pressure setting value (1, 6, 10), and a new TERUMO syringe (50 mL). The measured values may vary depending on the
test conditions.
* When the 1m extension tube (internal volume: 1.0 mL) is used.
Note
• The Occlusion alarm occurrence time with solution delivery at the minimum flow rate (0.01 mL/h) will be approx. 100 times of
those at the flow rate 1 mL/h.
Flow rate
(mL/h)
Occlusion
pressure
setting value
Occlusion pressure
Time to activate an
Occlusion alarm time
(min)
Bolus volume
(mL)
(kPa)
(kgf/cm
2
)
1
1
15
0.15 24.6 0.08
6
72
0.73 78.1 0.30
10
116
1.18 143.8 0.77
5
1
16
0.16 4.0 0.08
6
73
0.74 17.9 0.34
10
116
1.18 27.4 0.87