Specifications
Alarms
Occlusion alarm, Pressure alarm*
2
, Nearly Empty alarm, Slider Displacement
alarm, Syringe Barrel Detection alarm, Syringe Displacement alarm, Plunger
Displacement alarm, Battery alarm, Re-alarm, Shutdown Notice alarm, Power
Failure alarm, Start Reminder, No Flow Rate alarm, No VTBI alarm*
1
, Flow
Rate/VTBI Volume Judgment alarm*
1
, Completion alarm*
1
, Link Interruption
alarm*
2
.
*
1
: Only if the VTBI setting function is enabled (except the TCI mode) (Default:
disabled).
*
2
: Except the TCI mode
Safety functions
Remaining volume
detection setting
function
Sets the Nearly Empty alarm position based on the time before the syringe is
empty.
Occlusion detection
pressure select function
Selects the occlusion detection pressure setting values in 10 levels.
Bolus reduction
function
When an occlusion occurs, the slider is drawn back to reduce internal pressure.
Flow rate range setting
function A
(Soft limits)
The flow rate setting range can be controlled within the “flow rate setting range”
prescribed for each type of syringe (brand, size).
• When you try to set a value outside the range controlled by this function, a
message asking for confirmation is displayed.
If you confirm the setting, you can set the value which is outside the
controlled range and the display colour of the value changes.
Flow rate range setting
function B
(Hard limits)
The flow rate setting range can be controlled within the “flow rate setting range”
prescribed for each type of a syringe.
• Values outside the range controlled by this function cannot be set.
Keypad lock function
Disables switch operations. However, the Start switch, the keypad lock function
disable operation (press and hold the Back/Mute switch), and the Power switch
still function. When an alarm is activated, the keypad lock is released.
Available
Functions
Syringe installation
guidance function
Displays a screen that indicates the syringe installation status on the LCD.
Syringe brand display
function
Displays the syringe brand and size on the LCD.
Syringe type switch
function
Switching syringes is possible by performing the specified operation after the
power is turned on.
VTBI setting function
Sets the VTBI of dosage of drug solution.
VTBI time setting
function
Sets the VTBI time of dosage of drug solution.
Purge buzzer sound
function
When the Purge switch is being pressed, the buzzer sounds.
Bolus buzzer sound
function
During a bolus, the buzzer sounds.
Volume delivered clear
function (except the TCI
mode)
Clears the volume delivered display to zero.
Standby function
By pressing and holding the Back/Mute switch, the pump can be put in standby
mode. In the standby function, the Start Reminder is delayed.
Standby duration time
select function
Changes the time to activate a Start Reminder in the standby mode.
Buzzer volume select
function
Changes the buzzer volume.
Switch operation
buzzer sound function
Buzzer sounds when a switch operation is performed.