123
About External Communication Functions
Data Format
<Requesting basic operating status>
• Overview
The current basic operating status of the pump is obtained.
•
Command
: 0x01
•
SubCommand : 0x0000: Basic operating status is Request.
•
Data
Network system
Pump :
None
Pump
Network system : With data (See below.)
The following table lists the details of the “Basic operating status” data:
Basic operating status (Do not handle reserved data.)
Item
Size (Byte)
Details
Software Ver.
2
Range: 0x0000-0xFFFF
Basic operating status
1
Upper 4 bits
Lower 4 bits
0x0
Stop
(Reserved)
0x1
Start
(Reserved)
0x2
Standby
(Reserved)
0x3
Purge
(Reserved)
0x4
Hands On Bolus
(Reserved)
0x5
Hands Free Bolus
(Reserved)
0x6-0xF
(Reserved)
(Reserved)
(Reserved)
1
(Reserved)
(Reserved)
1
(Reserved)
Alarm information
3
b0 (low order)
Shutdown Notice alarm
b1
Link Interruption alarm
b2
Completion alarm
b3
Occlusion alarm
b4
Syringe Displacement alarm
b5
Plunger Displacement alarm
b6
Slider Displacement alarm
b7
Syringe Barrel Detection alarm
b8
Nearly Empty alarm
b9-b12
(Reserved)
b13
Power Failure alarm
b14
(Reserved)
b15
AC power detector function
b16
No Flow Rate alarm
b17
Flow Rate/VTBI Volume Judgment alarm
b18
No VTBI alarm
b19
Battery alarm
b20
Start Reminder
b21
Pressure alarm
b22
Maintenance timer
b23 (high order)
Battery Replacement Time/Failure alarm
One bit used for each item. (“1” indicates active.)