26
Preparation
Notes
• If the tube holder comes off, follow the procedure below to attach it.
• When inserting the notch of the tube holder into the hole on the rear side, insert while pulling the tube holder towards the rear
side. Forcibly inserting into the hole may cause damage to the notch.
1) Insert the notch of the tube
holder into the front side
hole on the pump.
2) Insert the notch of the tube
holder into the rear side hole
on the pump.
• The tube holder is firmly fixed to
the pump.
Check
Pole clamp
●
Clamp portion of
the pole clamp
●
Base portion of the pole clamp
●
Pole clamp lock portion
●
Pole clamp knob
1. There is no damage to any part of the pole clamp and that the clamp is clean.
2. When attaching pole clamp to the pump, it should lock in place firmly with an audible click.
3. There is no adhered drug solution on the pole clamp part and the clamp can be moved smoothly.
* If there is adhered drug solution, promptly clean in accordance with the directions on page 129.
Clean periodically even if there is no adhered drug solution.
2)
The following self-check (self-diagnosis) operations should be performed when the power is
turned on using the internal battery without a syringe installed.
1. The operation indicator lights in green and red alternately.
2. The message “Self-check” is displayed on the LCD, and the LCD flashes.
3. The buzzer sounds.
4. The alarm icons, slider displacement icon
, and syringe displacement icon
are flashing.
3)
Neither Battery alarm or Power Failure alarm has been issued.
1. The level of the battery icon on the LCD has two bars (green)
or more.
2. Neither Battery alarm (yellow) or Power Failure alarm (red) is displayed on the LCD.
4)
When the AC power cable is connected, the LCD displays AC icon
.
1. When the AC power cable is inserted into the AC inlet on the rear side of the main unit completely, the AC icon
is
displayed.
2. If the internal battery is not fully charged, fully charge it.
Note
• During charging,
and
flash alternately.
5)
When the AC power cable is connected, there is no looseness or any fault.
• The AC power cable is completely inserted into the AC inlet of the main unit.