About External Communication Functions
Item
Size (Byte)
Details
Unit of dilution
1
0x12:
μ
g/mL
0x13: mg/mL
Invalid value (when the function is disabled): 0x00
Drug library Ver.
2
Version of the drug library
Drug library is registered: 0x0001-0xFFFF
Drug library is not registered: 0x0000
Drug library number
2
Number of the drug library
Selecting dose mode (weight mode) No.1 (
μ
g/kg/min): 0x0FFD
Selecting dose mode (weight mode) No.2 (mg/kg/h): 0x0FFE
Selecting dose mode (weight mode) No.3 (mg/kg/h+DIPRIVAN): 0x0FFC
Invalid value (when the function is disabled): 0x0000
Occlusion pressure
setting
1
bit0-bit3: Setting of occlusion pressure. 1 to 10.
Occlusion pressure range: 0x01-0x0A
bit7: Setting of occlusion detection pressure 1 = In 10 levels
(bit4-bit6: Reserved)
Hands On Bolus flow
rate
2
100.00 to 1200.00 mL/h (100 mL/h step)
Hands Free Bolus VTBI 3
Value of 100 times the Bolus VTBI
Invalid value (when the function is disabled): 0x000000
Hands Free Bolus dose
time (Bolus VTBI time)
2
0 to 3600 s (1 s step)
Invalid value (when the function is disabled): 0x0000
Hands Free Bolus flow
rate
3
Value of 100 times the flow rate
Invalid value (when the function is disabled): 0x000000
Syringe brand number
1
Brand number of the syringe selected
OTCI library version
2
b0-b14 (lower)
Version number
Range: 0x01-0x7FFF
OTCI library unregistered: 0x0000
b15 (upper)
(Reserved)
OTCI library number
1
Range: 0x01-0x05
When the dose mode other than TCI is selected: 0x00
(Reserved)
1
(Reserved)
Prediction method
1
0x00: Plasma TCI, 0x01: Effect-site TCI
When the dose mode other than TCI is selected: 0x00
Age
1
12 to 100: 1 year step
Range: 0x0C-0x64
No age setting (the setting is cleared): 0xFF
Gender
1
0x00: Male, 0x01: Female
No gender setting (the setting cleared): 0xFF
Height
1
100-250 cm, 1 cm increments
Range: 0x00-0xFA
No height setting (the setting cleared): 0xFF
(Reserved)
1
(Reserved)
Target concentration
2
Value of 100 times the concentration 0.00 to 50.00 (0.10 step)
Range: 0x0000-0x1388
When the dose mode other than TCI is selected: 0x0000
Target concentration
unit
1
0x11: ng/mL, 0x12:
μ
g/mL
When the dose mode other than TCI is selected: 0x00
Decrement
concentration
2
Value of 100 times the concentration 0.01 to 50.00 (0.10 step)
Range: 0x0001-0x1388
When the dose mode other than TCI is selected: 0x0000