22
Precautions
Cautions
• Regular inspections should be conducted. If any fault is observed, stop using this product and request an inspection and
repair. [The original functionality or performance of this product may not be achieved.]
• Attention is required not to apply any strong static electricity. [It may result in failure or malfunction.]
• Prior to first use, or if unused for a long period, connect this product to an earthed AC power supply and provide a
sufficient charge (8 hours or more) with the power turned off. [If not sufficiently charged, this product may not be able to
operate using the internal battery during a power failure, etc.]
• Even within the normal conditions of use, this product should not be used under conditions that cause a sudden
temperature change. [Condensation inside the product results in damage and time degradation, and therefore the original
functionality or performance of this product may not be achieved.]
• The AC power cable used for this product should not be pinched with forceps or equipment, e.g. caster, or punctured with
a needle. [If a cable is damaged, electric shock or fire may occur. The original functionality or performance of this product
may not be achieved.]
• This product should not be used in a place where vibration, dust, mist, corrosive gas, etc., occurs or in a place where
the product is sprayed with liquid. If solution is spilled on this product, thoroughly wipe with a dry soft cloth. [The original
functionality or performance of this product may not be achieved, and it may cause a malfunction.]
• In a case where a manufacturer has made changes to their syringe, the flow rate accuracy and alarm function may not be
guaranteed. If any fault is observed, immediately stop using the product and contact TERUMO trained service technicians.
• Avoid any location for placement that may cause interference with other infusion pumps.
• When transporting with the pump fixed to the IV pole, do not hold the handle of the pump nor apply pressure from the top.
[The pole clamp may be dislocated or damaged.]
• When this product is transported while in use, do not touch the switches, etc. Use the keypad lock function as needed.
[Touching the switches etc. without the keypad lock enabled may result in an operation (power on/off, stop, start, rapid
infusion) that is not intended by this product.]
• The LCD and the Operation panel (switches etc.) of this product should not be pressed with excessive force and should not
be operated with a pointed object. [It may cause damage to or failure of the LCD or the Operation panel.]
• Do not disassemble, make alterations (including actions that interfere with the functionality or performance such as
taping the LCD or a movable part) to, or repair this product. [This may result in failure, damage or device performance
degradation of this product.]
• This product should only be used with a sufficient power supply. [If the power supply is not sufficient, the internal battery
will be used for the operation, which may result in no power supply in an emergency situation.]
• When using devices (mobile phones, radio devices, radio knives, defibrillators, etc.) that emit an electromagnetic wave
within the area of this product, they should be used as far away as possible and you should verify the normal operation of
this product in the configuration it is used. In addition, this product should use a power supply of a system isolated from
these devices and should obtain secure earth connection. [Malfunction in this product may result in critical harm to the
patient.]
• When used in the vicinity of a radio knife (Medical radio knives are surgical
equipment for incision and coagulation by high energy radiofrequency
current), check the following before use.
(1) Radio knives have a different level of high-frequency noise emission
depending on the types, and using old models (vacuum tube gap type)
in combination should be particularly avoided as the noise levels from
those are higher.
(2) The distance from the radio knife cord (knife holder, knife cord and
return electrode cord) and radio knife body to this product should be kept
to a minimum of 50 cm.
(3) The radio knife and this product should be operated by a power supply
from a separate system, and both should be securely earthed.
• When connecting this product with other medical equipment or network system, check its conformance to IEC 60601-
1:2005+A1:2012 (EN 60601-1:2006+A1:2013), IEC 60601-1-2:2007 (EN 60601-1-2:2007) and IEC 60601-1-2:2014 (EN
60601-1-2:2015) prior to use in order to ensure the system safety.
• When connecting this product with a network system or other equipment, for safety, checking the manufacturer’s
specifications of the equipment is recommended. In addition, use EMI compliant products for the connection cable.
• When using the external communication function, pay particular attention, as it is more susceptible to the effect from a
radio knife, mobile phone, radio device, defibrillator, etc. Regularly check that this product is operating normally.
Medical
electrical
equipment
Power supply for
medical electrical
equipment
Keep 50 cm or more
Power supply from a different system
Power supply
for radio knife
Knife holder
Return electrode
Radio knife