53
Operation Procedure
Confirmation
switch
7)
Press the Confirmation switch.
Check
• The syringe brand and the syringe size are displayed on the LCD correctly.
Note
• If the flow rate screen appears, go to step 11).
Back/Mute switch
8)
Press the Back/Mute switch.
Display select switch
Dial
9)
Turn the Dial or press the Display select switch to select
the dose mode.
Notes
• This section shows an example of selecting mL/h for the dose mode.
For other dose modes, see pages 66 and 75.
• Turning the Dial or pressing the Display select switch moves the white frame.
Confirmation
switch
10) Press the Confirmation switch.
Note
• When the Confirmation switch is pressed, the flow rate display appears.