65
Other Operation Procedures
Standby Function
The standby function can extend the Start Reminder.
Check
• The syringe is set and the solution delivery is ready to be started.
Back/Mute
switch
2 seconds
or more
•
Press and hold the Back/Mute switch for 2 seconds or more.
(The standby mode is entered.)
Note
• The LCD displays the message “Stand By”.
Notes
• The default standby time is 24 hours. To change this, contact TERUMO trained service technicians (the time can be changed to
24 hours or less).
• Turning the Dial or pressing any switch cancels standby mode.
• The standby mode cannot be entered when an alarm has been issued. Eliminate the cause of the alarm first.