62
Operation Procedure
Replacing Syringe to Continue Solution Delivery
Stop switch
Operation indicator
1)
Press the Stop switch to stop the solution delivery.
Note
• For details on stopping solution delivery, see page 61.
2)
Close the infusion line.
Note
• For details on stopping solution delivery, see page 61.
3)
Remove the syringe.
Note
• For details on removing the syringe, see page 61.
4)
Check that the syringe brand to be used matches the
syringe brand displayed on the LCD, and then place the
syringe.
Note
• For details on setting the syringe, see page 51.
5)
Perform priming before starting solution delivery to the
patient, and keep the plunger pushed by the slider.
Note
• For details on priming, see page 58.
6)
Clear the volume delivered as required.
Note
• For details on clearing the volume delivered, see page 64.
Start switch
Operation indicator (green flashing)
7)
Switch the solution delivery to the patient side, open
the infusion line, and then press the Start switch to start
solution delivery.
Check
• The flow rate is correctly set.
Notes
• The operation indicator flashes green.
• The volume delivered increases.