60
Operation Procedure
Starting Solution Delivery
Start switch
Operation indicator (green flashing)
•
Check the flow rate setting and press the Start switch to
start solution delivery.
Notes
• The buzzer sounds.
• The operation indicator flashes green.
• The volume delivered increases.
• Flow rate can be changed during solution delivery. Rotate the Dial in the flow rate
display to change the flow rate and the confirmation message appears. Press the
Start switch to start solution delivery at the changed flow rate.
Volume delivered
Notes
• When the pump is fully set but not started, after 2 minutes the LCD displays the
Start Reminder screen (yellow) and the buzzer sounds.
• Use the standby function (see page 65) if you want to temporarily extend the
Start Reminder.
When Solution Delivery Cannot Start
The alarm sounds and the following warnings display. (See page 133.)
No Flow Rate alarm (yellow)
The flow rate has not been set. Set the flow rate and press the Start switch again.
Cautions
• If soft and hard limits are not activated before starting solution delivery, check the settings for any mistakes such as
an error in the digit of flow rate. This product does not have a function to determine whether or not the set values are
appropriate, possibly resulting in excessive infusion or insufficient infusion to a patient.
• If any fault is observed, immediately stop the solution delivery and check the setting and the attachment status of the
syringe.