46
Operation Procedure
4)
Remove the syringe.
Note
• For details on removing the syringe, see page 61.
5)
Check that the syringe brand to be used matches the
syringe brand displayed on the LCD, and then place the
syringe.
Warning
• Before using the TCI mode, make sure the correct drug name, model, and
target concentration are displayed. Incorrect settings of those may cause
significant health hazard to the patient.
Note
• For details on setting the syringe, see page 32.
6)
Perform priming before starting solution delivery to the
patient, and keep the plunger pushed by the slider.
Note
• For details on priming, see page “Priming” on page 38.
7)
Switch the solution delivery to the patient side, and open the
infusion line.
Start switch
Operation indicator (green flashing)
8)
Press the Start switch.
Check
• The setting of target concentration (Cpt or Cet) should be correct.
Notes
• The buzzer sounds.
• The operation indicator flashes green.
Warning
• When removing the syringe installed in this product, first close the three way stopcock etc. of the infusion line. Otherwise
excessive infusion (siphoning free flow by gravity) of the drug solution may occur.