Changing Settings on the Menu
Plasma limit setting (TCI mode only)
Before starting solution delivery in effect-site TCI, set the upper limit of plasma concentration for the appropriate level of overshoot.
Display
select switch
2 seconds
or more
1)
Press and hold the Display select switch for 2 seconds or
more.
Confirmation
switch
2)
Press the Confirmation switch.
Note
• The current setting display turns from light blue to dark blue (selected status).
Display select switch
Dial
3)
Turn the Dial or press the Display select switch to select the
‒
TCI Menu.
Note
• Turning the Dial or pressing the Display select switch moves the white frame.
Confirmation
switch
4)
Press the Confirmation switch.
Notes
• The TCI Menu screen appears.
• The TCI Menu screen can also be displayed if you turn the Dial or press the
Display select switch in step 2) to select the TCI tab.