91
Changing Settings on the Menu
10) The bolus ends when the VTBI is reached.
Check
• The screen returns to the flow rate screen when the VTBI has been reached and
the bolus has been completed.
Notes
• The volume delivered display area displays the amount of bolus during bolus
infusion, and displays the sum of the amount of bolus and the volume delivered
when the VTBI is attained and the bolus ends.
• When bolus infusion during solution delivery is completed, solution delivery
continues at the flow rate before the bolus infusion.
• A Completion alarm occurs if the volume delivered reaches the VTBI when bolus
infusion is used during solution delivery. The solution delivery continues at the flow
rate of KVO (Keep Vein Open).
• Pressing the Stop switch forcibly stops Hands Free Bolus.