93
Changing Settings on the Menu
Confirmation
switch
5)
Press the Confirmation switch.
Increases value
Dial
Decreases
value
6)
Turn the Dial to select the maximum flow rate.
Note
• The upper limit of flow rate can be set within the range of the following values.
Setting range:
0.01 to 999.00 mg/kg/h (Propofol)
0.01 to 999.00
μ
g/kg/min (Remifentanil)
Step:
0.01 step (0.01 to 10.00)
0.10 step (10.00 to 100.00)
1.00 step (100.00 to 999.00)
Confirmation
switch
7)
Press the Confirmation switch.
Notes
• The display returns to the graph screen.
• When the maximum flow rate is changed before starting solution delivery, the
indications of the initial infusion dose/time and the prediction lines in the graph will
change accordingly.