LISST-
200X User’s Manual
33
LISST-200X Operation Details
•
The Burst and Fixed sample rate modes are used to set up
a sampling program that will save data onboard the
instrument. The icons next to the various values give a
better understanding of their meaning. The software
automatically checks the values entered to make sure that
there is no conflict. For example, when a Samples per
Average value is entered, the minimum sample interval is
computed. If this value is less than the minimum permitted
the value will be changed to the minimum and the text will
turn red. A similar test is done on the Burst Interval setting.
4
•
Select the Start Conditions tab to configure when the
instrument will begin sampling.
•
There are five options: Depth, Time, External Mechanical
Switch, External Digital Input, and Time Delay. Select the
mode by clicking on the button next to the its label. Select
the correct parameters as required.
Start Condition
Set
•
If the AutoStart check box is selected, the LISST-200X will
start the sampling program when power is applied to the
instrument. If this is not selected the user must send the